Model Number WA90004W
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bowel Perforation (2668)
Type of Reportable Event Death
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient's outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that after a therapeutic colonoscopy procedure at an unknown date, where the colon and the transversum were treated using smart argon, the procedure was first believed to have been finished normally, when the patient's condition deteriorated and a follow-up procedure was planned.However, the patient passed away before it could be carried out.There were no reports of any device malfunctions during the initial procedure.
 
Additional Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation but the olympus regional repair center (rrc) hamburg, germany (received at the rrc on 2021-05-06).The evaluation at the rrc did not reveal any malfunctions of the argon plasma coagulation unit or the concomitant esg-300 hf generator but found both devices to be working correctly.Furthermore, there were no reports of any malfunction of any of the olympus medical devices used during the initial procedure.Based on the information available, there is no indication for a technical defect that might have contributed to the patient¿s death.The exact cause of the reported phenomenon and the patient¿s outcome could not be determined and is being judged as unknown.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected serial numbers of both the argon plasma coagulation unit and the concomitant hf generator without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARGON PLASMA COAGULATION UNIT "APU-300"
Common Device Name
ELECTROSURGICAL UNITS, MONOPOLAR, ARGON-ENHANCED COAGULATION
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key11789035
Report Number9610773-2021-00131
Device Sequence Number3921504
Product Code GEI
UDI-Device Identifier04042761083560
UDI-Public04042761083560
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K180200
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberWA90004W
Device Catalogue NumberWA90004W
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/06/2021
Supplement Date Received by Manufacturer06/07/2021
Initial Report FDA Received Date05/07/2021
Supplement Report FDA Received Date06/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS ELECTROSURGICAL GENERATOR "ESG-300"; OLYMPUS ELECTROSURGICAL GENERATOR "ESG-300".
Outcome Attributed to Adverse Event Death;
-
-