• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA ONE 3.1
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2021
Event Type  malfunction  
Manufacturer Narrative
Unique identifier (udi) #: (b)(4).A full analysis of the data logs has been performed and this analysis concluded that the arm connection failed due to a memory allocation issue that prevented its complete initialization.Consequently, the software asked to restart the application or shut down the computer.In order to solve the issue, the user had to shut down and reboot the computer.This issue experienced will be addressed through the continuous improvement of our product.
 
Event Description
After turning on the robot for an seeg, the emergency stop button was lit up but not pressed.The robot was restarted and attempted to connect but failed.The software was restarted and still failed to connect.The robot was completely turned off and unplugged.After rebooting, the robot successfully connected.This caused a 5 minutes delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA ONE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key11789039
MDR Text Key258703868
Report Number3009185973-2021-00118
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200511
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROSA ONE 3.1
Device Catalogue NumberROSAS00203
Device Lot Number3.1.4.1650
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-