ZIMMER BIOMET, INC. LOCK SCREW SQ 2.7MM 18MM STE; PLATE, FIXATION Back to Search Results
Model Number 131827118
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Foreign- (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the incoming inspection team member found debris in the sterile package.There was no patient or surgical involvement at the time of the event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d9, g3, g6, h1, h2, h6, h10.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product/provided pictures found hair-like debris in the sterile packaging.Device history record (dhr) was reviewed and no discrepancies were found.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCK SCREW SQ 2.7MM 18MM STE
Common Device Name
PLATE, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11789752
Report Number0001825034-2021-01230
Device Sequence Number11742632
Product Code HRS
UDI-Device Identifier00887868450297
UDI-Public(01)00887868450297(17)301012(10)983620
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112345
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number131827118
Device Catalogue Number131827118
Device Lot Number983620
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/21/2021
Supplement Date Received by Manufacturer06/03/2021
Initial Report FDA Received Date05/07/2021
Supplement Report FDA Received Date06/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-