Model Number 3243
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 04/28/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the burr likely contributed to the guidewire fracture.The 95% stenosed target lesion was located in the moderately tortuous and severely calcified proximal left anterior descending (lad) artery.A 1.75mm rotapro and a rotawire and wireclip torquer were selected for a percutaneous coronary intervention.A non-boston scientific guidewire was advanced through the lesion in the lad and into the distal lad.The non-boston scientific guidewire was exchanged to a rotawire using a non-boston scientific microcatheter.The 1.75mm rotapro was advanced over the rotawire and rotablation was performed on the proximal lad.The physician then decided to rotablate distally in the mid lad but with a smaller 1.5mm burr.The 1.75mm rotapro was withdrawn over the rotawire using dynaglide.A large kink was noted in the proximal end of the wire.During removal of the rotawire, a fracture occurred.An opticross catheter visualized that approximately 2cm of the guidewire tip remained entirely in the small septal artery and as such could not be snared.The remaining rotawire was withdrawn.The procedure was continued with a new rotawire and a 1.5mm rotablator.Rotablation was performed in the mid lad region with success.The 1.5mm rotablator was withdrawn and the lad was stented with a des stent and post dilated.The patient fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11790280
Report Number2134265-2021-05884
Device Sequence Number19820664
Product Code MCX
UDI-Device Identifier08714729893370
UDI-Public08714729893370
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P900056
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model Number3243
Device Catalogue Number3243
Device Lot Number0026107269
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/05/2021
Initial Report FDA Received Date05/07/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE: BMW; MICROCATHETER: FINECROSS
Outcome Attributed to Adverse Event Required Intervention;
-
-