| Model Number |
72202088S |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident. a visual inspection found distal tip damage, fiber damage, and a cracked distal lens.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution. a complaint history review concluded this was a repeat issue.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during the surgery the scope had scratched lens.No patient injuries or delay reported.Smith and nephew back-up device was available to complete the surgery.Results of investigation have concluded that this unit had a cracked distal lens which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|