Model Number 1549-07-001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 04/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address poly spin out due to the gap between the femur and tibia being too large.Surgeon implanted a hinge with thick tray to close the joint space.No surgical delay has been reported.Doi: (b)(6) 2020 dor: (b)(6) 2021 left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE POS FEM AUG SZ 7 4MM
Common Device Name
KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key11791154
Report Number1818910-2021-09777
Device Sequence Number14414653
Product Code JWH
UDI-Device Identifier10603295383093
UDI-Public10603295383093
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1549-07-001
Device Catalogue Number154907001
Device Lot NumberJ10W58
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/22/2021
Supplement Date Received by Manufacturer06/14/2021
Initial Report FDA Received Date05/07/2021
Supplement Report FDA Received Date06/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CRS FEMORAL LT SZ 7 CEM; ATTUNE CRS RP INSRT SZ 7 26MM; ATTUNE DIST FEM AUG SZ 7 12MM; ATTUNE DIST FEM AUG SZ 7 12MM; ATTUNE POS FEM AUG SZ 7 4MM; ATTUNE POS FEM AUG SZ 7 4MM; ATUN FEM SLV M/L 40MM FULL POR; ATUN PRESSFIT STR STEM14X110MM; ATUN TIB SLV M/L 37MM FULL POR; ATUNE PRESSFIT STR STEM12X60MM; ATUNE REV RP TIB BASE SZ 5 CEM; ATTUNE CRS FEMORAL LT SZ 7 CEM; ATTUNE CRS RP INSRT SZ 7 26MM; ATTUNE DIST FEM AUG SZ 7 12MM; ATTUNE DIST FEM AUG SZ 7 12MM; ATTUNE POS FEM AUG SZ 7 4MM; ATTUNE POS FEM AUG SZ 7 4MM; ATUN FEM SLV M/L 40MM FULL POR; ATUN PRESSFIT STR STEM14X110MM; ATUN TIB SLV M/L 37MM FULL POR; ATUNE PRESSFIT STR STEM12X60MM; ATUNE REV RP TIB BASE SZ 5 CEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
-
-