MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD A-LINE; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD A-LINE; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 364356
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: bd received 12 boxes of product for investigation.The samples were evaluated by visual examination and the indicated failure mode for torn blister pack with the incident lot was observed in 2 blisters.Additionally, 10 retention samples from bd inventory were evaluated by visual examination and no packaging damage was observed.Bd was not able to determine when or where the tear to the packaging could have occurred.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.
 
Event Description
It was reported the bd a-line¿ devices were damaged or had an open unit package/seal issue where sterility was compromised.This event occurred 12 times.The following information was provided by the initial reporter.The customer reported "some packages were torn.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD A-LINE
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key11791870
MDR Text Key249964457
Report Number9617032-2021-00418
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K022426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/31/2022
Device Catalogue Number364356
Device Lot Number0239030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-