• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US CONE REAMER EXTENSION; HIP INSTRUMENTS : ADAPTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US CONE REAMER EXTENSION; HIP INSTRUMENTS : ADAPTORS Back to Search Results
Model Number 2576-00-008
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported as "instruments were used in surgery.Upon inspection of those instruments this morning, (b)(6) , we noticed some defects in the two sleeve trials making it difficult to use with the "sleeve introducer" handle.We also inspected the reamer handle and it was difficult to lock on the conical reamers.These issues were not mentioned in surgery, however, after speaking with the surgical tech today, (b)(6) , he asked me to take a look." no surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.Corrected: h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that the handle sheath was difficult to fully pull down, perhaps, needed to be lubricated.The sleeves had appeared to be cut by a blade or osteotome around the top making it difficult for the introducer to work.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONE REAMER EXTENSION
Common Device Name
HIP INSTRUMENTS : ADAPTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11794033
Report Number1818910-2021-09839
Device Sequence Number14460107
Product Code HTO
UDI-Device Identifier10603295145301
UDI-Public10603295145301
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2576-00-008
Device Catalogue Number257600008
Device Lot NumberSO2013062
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2021
Initial Date Received by Manufacturer 04/29/2021
Supplement Date Received by Manufacturer05/18/2021
06/15/2021
Initial Report FDA Received Date05/07/2021
Supplement Report FDA Received Date05/27/2021
06/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age58 YR
-
-