Model Number 91496 |
Device Problem
Device Alarm System (1012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Spacelabs has launched an investigation and will file a supplemental report when the investigation is complete.
|
|
Event Description
|
Spacelabs received a report on (b)(6) 2021 customer reported the command module failed to alarm for pause.No injury was reported with this event.
|
|
Manufacturer Narrative
|
No failure was found.The device historical data and device logs recognized a pause event with no corresponding alarm.It could not be determined that a malfunction occurred during this complaint episode.If additional information is provided, spacelabs will reassess the investigation.This investigation is considered complete and closed.
|
|