| Model Number |
67SP040 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cuffs are not inflating on smiths medical trach tubes and seeming to stick to one side.Takes a lot of manipulation and sterile water to inflate the cuff.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: investigation completed on a smiths medical tracheostomy|silicone - bivona tubes neo/ped received a sample from two p/n 67sp040 l/n 3763982 and p/n 67sn035, l/n 3949041 were testing following ifu instruction.The pressures : 44.11% and sample 2:42.10%.These results are over 33.3% that is the minimum acceptable.Based on the test, the units are within spec.The complaint is not confirmed.The complaint was not confirmed.
|
| |
|
Event or Problem Description
|
|
Investigation completed and summary in h 10.
|
| |
|