PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR Back to Search Results
Model Number LC580IM
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been made available for evaluation at this time.Should the device or further information become available, a follow-up report will then be issued.
 
Event or Problem Description
Alleges the back broke when consumer was in the process of raising the chair to get out of it.
 
Additional Manufacturer Narrative
The returned lift chair was subjected to extreme abuse/misuse.The chair's articulation had nothing to do with the back breaking.The seat back appeared to have been stapled and glued correctly.The abuse/misuse included but was not limited to liquid ingress, operating against an obstruction and excessive force on the back.This pattern of abuse likely led to the back breaking.
 
Event or Problem Description
Alleges the back broke when consumer was in the process of raising the chair to get out of it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
ELECTRIC POSITIONING CHAIR
MDR Report Key11796069
Report Number2530130-2021-00039
Device Sequence Number3847142
Product Code INO
UDI-Device Identifier00606509400770
UDI-Public00606509400770
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K091578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberLC580IM
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2021
Supplement Date Received by Manufacturer04/27/2021
Initial Report FDA Received Date05/10/2021
Supplement Report FDA Received Date09/27/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/21/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-