MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI CERV SPINE LOCKING PLATE 24MM (16MM); APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI CERV SPINE LOCKING PLATE 24MM (16MM); APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Model Number 450.116
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 04/13/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Part # 450.116-us, lot # 23p9644, manufacturing site: (b)(4), release to warehouse date: december 11, 2019.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.The implant(s) was not returned and instead the investigation will be done based on the supplied image(s).The image(s) was reviewed, and the complaint condition could be confirmed as the x-ray shows a broken locking plate.As the implant(s) was not returned in as received condition, dimensional inspection, material or drawing reviews are not applicable.A definitive assignable root cause could not be determined based on the provided information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, the patient underwent a revision surgery for a broken plate.On (b)(6) 2020, due to a traumatic injury, the patient had an anterior cervical fusion at c5-c6 using a cslp plate.During a routine follow-up visit on (b)(6) 2021, the surgeon discovered that the plate had broken when viewing on a standard follow-up x-ray.The surgeon performed the revision surgery on (b)(6) 2021.The surgeon confirmed that the patient has a non-union at the c5-c6 level.During the revision, he removed the broken cslp plate.The expansion head screws and locking screws were also completely removed and no fragments/parts remain in the patient.One the entire cslp construct was removed.The surgeon flipped that patient prone and did a c5-c6 posterior cervical fusion using the depuy synthes spine symphony system.The procedure was successfully completed.This report involves one (1) ti cervical spine locking plate 24mm (16mm).This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI CERV SPINE LOCKING PLATE 24MM (16MM)
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11797332
Report Number2939274-2021-02273
Device Sequence Number14618207
Product Code KWQ
UDI-Device Identifier10705034780504
UDI-Public(01)10705034780504
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K031276
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number450.116
Device Catalogue Number450.116
Device Lot Number23P9644
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/13/2021
Initial Report FDA Received Date05/10/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1.8MM TI LOCKING SCREW; 1.8MM TI LOCKING SCREW; 1.8MM TI LOCKING SCREW; 1.8MM TI LOCKING SCREW; 4.35MM TI CANC EXP HD SCR/SLF-TAP/16MM; 4.35MM TI CANC EXP HD SCR/SLF-TAP/16MM; 4.35MM TI CANC EXP HD SCR/SLF-TAP/16MM; 4.35MM TI CANC EXP HD SCR/SLF-TAP/16MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age22 YR
-
-