| Model Number |
24701 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
04/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter and guidewire entrapment, pain and aborted procedure occurred.The 100% stenosed target lesion was located in the mildly calcified distal superficial femoral artery.A 3.5mm x 40mm x 145cm coyote es balloon catheter was advanced for dilatation.The balloon was deployed; however, during withdrawal, the physician could not remove the balloon over the choice guidewire.The balloon and the wire were entirely removed as one unit.The physician decided to abort the procedure because the patient was having so much back pain.No patient complications were reported.The patient will return at a later date to be treated with atherectomy.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a coyote es balloon catheter with a 0.014 guidewire stuck inside.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual and microscopic examination revealed no damages.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter and guidewire entrapment, pain and aborted procedure occurred.The 100% stenosed target lesion was located in the mildly calcified distal superficial femoral artery.A 3.5mm x 40mm x 145cm coyote es balloon catheter was advanced for dilatation.The balloon was deployed; however, during withdrawal, the physician could not remove the balloon over the choice guidewire.The balloon and the wire were entirely removed as one unit.The physician decided to abort the procedure because the patient was having so much back pain.No patient complications were reported.The patient will return at a later date to be treated with atherectomy.
|
| |
|
Search Alerts/Recalls
|