COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number E7510-25
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Full thickness (Third Degree) Burn (2696)
Date of Event 04/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, post-operatively on inguinal hernia surgery, the device burnt the patient's body.The new pad was applied to clean dry skin but the back side of the patient was wet because of cream.The patient was in pediatric intensive care unit due to low body temperature and found the burn four hours later.Photograph was received and showed the patient's buttocks was burn with blisters.
 
Additional Manufacturer Narrative
Additional information: g1 (mfr name, mfr contact first name, last name, email, phone number), g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the patient had red marks and the pad was not shown in the photo.Functional testing found the pad was not shown.It was reported that use of the device resulted in a burn.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: slowly remove the return electrode with one hand while supporting the skin with the other hand to avoid skin trauma.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key11797877
Report Number1717344-2021-00690
Device Sequence Number3868891
Product Code GEI
UDI-Device Identifier10884524001043
UDI-Public10884524001043
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K822572
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/21/2022
Device Model NumberE7510-25
Device Catalogue NumberE7510-25
Device Lot Number02470111X
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/19/2021
Supplement Date Received by Manufacturer07/23/2021
Initial Report FDA Received Date05/10/2021
Supplement Report FDA Received Date07/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age2 MO
-
-