• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN OFFSET COUPLER TRIAL 2MM; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN OFFSET COUPLER TRIAL 2MM; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71434302
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a tka, a lgn offset coupler trial 2mm did not function correctly.Instrument outside the patient.The procedure was completed without delay using a s+n back-up device.Patient was not harmed.
 
Additional Manufacturer Narrative
H6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device could not confirm the stated failure mode but could confirm that the device has scratches on it.A functional evaluation was performed on the device and confirmed the stated failure mode.The internal adjustment screw inside of the coupler will no longer tighten rendering the device inoperable.The device shows signs of extensive use a complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN OFFSET COUPLER TRIAL 2MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11798469
Report Number1020279-2021-03922
Device Sequence Number19603341
Product Code HRY
UDI-Device Identifier03596010547965
UDI-Public03596010547965
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71434302
Device Catalogue Number71434302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2021
Initial Date Received by Manufacturer 04/15/2021
Supplement Date Received by Manufacturer06/30/2021
Initial Report FDA Received Date05/10/2021
Supplement Report FDA Received Date08/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-