Model Number 74018911
Medical Device Problem Code Incomplete or Inadequate Connection (4037)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a tka, a journey art isrt assm tool did not work properly to help seat poly into baseplate.Instrument outside the patient.The procedure was completed without delay using a s+n back-up device.Patient was not harmed.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device shows signs of extensive use.A functional evaluation was performed on the device and confirmed the stated failure mode.The locking mechanism of the returned device has seized rendering the device inoperable.A review of complaint history did not reveal additional complaints for the listed item.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY ART ISRT ASSM TOOL
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11798554
Report Number1020279-2021-03923
Device Sequence Number3081473
Product Code HRY
UDI-Device Identifier03596010557292
UDI-Public03596010557292
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74018911
Device Catalogue Number74018911
Device Lot Number15APG0019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2021
Initial Date Received by Manufacturer 04/15/2021
Supplement Date Received by Manufacturer05/22/2021
Initial Report FDA Received Date05/10/2021
Supplement Report FDA Received Date05/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-