| Model Number |
6937A-52 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: d334drg icd implanted: (b)(6) 2013; 5076-45 lead implanted: (b)(6) 2013.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that a remote monitoring transmission was received and was reviewed.Alerts were noted to have been triggered for hi gh/undefined superior vena cava coil impedance on the right ventricular (rv) lead as well as for high/undefined atrial impedance a few years earlier.The svc coil was noted to have been programmed off and the device was noted to be in vvi mode.The leads remain in use.No patient complications have been reported as a result of this event.
|
| |
|