• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 102956
Device Problem Device Difficult to Setup or Prepare (1487)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/22/2021
Event Type  malfunction  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Event Description
It was reported that the centrimag motor could not connect, so it was sent back to abbott and a new motor was provided.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of the centrimag motor not being able to connect to the centrimag system was not confirmed.The returned centrimag motor (serial number: (b)(6)) was functionally tested at the european distribution center.The unit was tested with known working test centrimag equipment.The reported event was not reproduced as the motor functioned as intended during testing without any issues or atypical alarms being produced.Preventative maintenance was performed, and the serviced and tested unit was returned to the customer after passing all tests per procedure.The root cause of the reported event could not be conclusively determined through this analysis.The device history records were reviewed and the records revealed that the centrimag motor, serial number (b)(6), was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 10 ¿ ¿emergencies/troubleshooting¿ provides instructions for operation when there is a need to exchange the main console or motor with a backup console or motor.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.Centrimag motor instructions for use instructs the user to inspect the centrimag motor, cable, console connector, and locking mechanism for any damage prior to use.If any component is damaged, do not use the centrimag motor.This document states that if the unit fails to operate according to the motor specifications or a console diagnostic error indicates a centrimag motor malfunction, it should be returned.Additionally, this document instructs the user to always have a spare centrimag motor and back-up equipment available.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, US
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
MDR Report Key11801039
MDR Text Key251243671
Report Number3003306248-2021-01106
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number102956
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2021
Was the Report Sent to FDA? No
Date Manufacturer Received07/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-