• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES ARCHITECT STAT MYOGLOBIN REAGENT KIT; MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES ARCHITECT STAT MYOGLOBIN REAGENT KIT; MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL Back to Search Results
Model Number 2K43-20
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient identifier = sid= (b)(6).(b)(4).Was this device serviced by a third party?: no.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed falsely elevated architect stat myoglobin results on one patient.The results provided were: sid (b)(6) initial=1003.7 ng/ml / repeated due to questioned by physician =51 ng/ml there was no reported impact to patient management.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The complaint evaluation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history record review and in-house testing of architect stat myoglobin reagent lot 57557un20.Trending review determined no adverse trend for the issue for the product.Return testing was not completed as returns were not available.Trending review determined no adverse trend for the issue for the product.Accuracy testing was performed using a retained kit of lot 57557un20 with panels, which mimic patient samples, and all specifications were met indicating that the lot is performing acceptably.Device history record review did not identify any non-conformances or deviations with the complaint lot and complaint issue.A review of the manufacturing documentation did not identify any issues associated with the customer¿s observation.There was no issue with reagent performance identified.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation no product deficiency was identified for the architect stat myoglobin reagent lot 57557un20.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT STAT MYOGLOBIN REAGENT KIT
Common Device Name
MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064
MDR Report Key11801795
Report Number1415939-2021-00022
Device Sequence Number11613431
Product Code DDR
UDI-Device Identifier00380740003296
UDI-Public00380740003296
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model Number2K43-20
Device Catalogue Number02K43-20
Device Lot Number57557UN20
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/27/2021
Supplement Date Received by Manufacturer06/21/2021
07/23/2021
Initial Report FDA Received Date05/10/2021
Supplement Report FDA Received Date07/01/2021
07/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I1000SR MOD, 01L86-01, (B)(6).
-
-