• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: 5.0FR URETHANE UMB CATH; CATHETER, UMBILICAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

5.0FR URETHANE UMB CATH; CATHETER, UMBILICAL ARTERY Back to Search Results
Model Number 8888160341
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.  if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.  as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event or Problem Description
Customer reports: the single lumen umbilical arterial catheter was leaking blood.It was noted that there was a hole in the catheter tubing within the first 2 inches from connection end.
 
Additional Manufacturer Narrative
A device history record (dhr) review revealed no discrepancies that may have contributed to a complaint of this reported issue.All quality assurance testing performed during manufacturing was acceptable.The quality assurance review of the visual, physical, and dimensional evaluation results concludes the product met specification requirements.In addition, all dhr were reviewed prior to the product release to confirm all requirements were met.One used catheter from lot number 1923400325 and product code 8888160341 was received inside of its original package opened for analysis and investigation.The visual inspection revealed that the catheter was used due to residues of blood that was identified.The catheter showed an extension that does not belong to the original product.During the physical evaluation the sample was submitted to underwater testing and a leak was identified.The reported event was confirmed.Based on the event description "chlorhexidine 2 and 70% alcohol were used to clean the device before insertion, but it was not used on the catheter tube." according to the instructions for use (ifu) of the product, ¿do not use alcohol, acetone, or alcohol containing antiseptics directly on the catheter.Carefully check antiseptic solutions for alcohol or acetone.These substances may cause irreversible damage to the polyurethane which can lead to a leak or break¿.Therefore, the most probable root cause can be considered as a deviation from the ifu.According to occurrence analysis performed for this complaint the probability calculated of occurrence of harm (p1 x p2) is (o-1) is equal than the probability expected of occurrence of harm (p1 x p2) is (o-1) defined in the risk analysis chart (rac) for the argyle¿ umbilical vessel catheters (uvcs)¿.Therefore, a corrective or preventive action is not deemed necessary.A formal corrective/preventative action (capa) was not required for this issue and according to procedure cardinal health distributed product field action assessment (dpfa) procedure¿ a dpfa was not required for this issue.It must be noted that in-process controls such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling are performed at the plant t to prevent non-conforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0FR URETHANE UMB CATH
Common Device Name
CATHETER, UMBILICAL ARTERY
MDR Report Key11802627
Report Number3009211636-2021-00746
Device Sequence Number4492081
Product Code FOS
UDI-Device Identifier20884527005120
UDI-Public20884527005120
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8888160341
Device Catalogue Number8888160341
Device Lot Number1923400325
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 05/08/2021
Supplement Date Received by Manufacturer05/08/2021
Initial Report FDA Received Date05/10/2021
Supplement Report FDA Received Date09/01/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device N
Patient Sequence Number1
-
-