HOLOGIC, INC AFFIRM PRONE BIOPSY SYSTEM, 3D; BREAST BIOPSY SYSTEM Back to Search Results
Model Number PBX-SYS-AFFIRM-3D
Medical Device Problem Code Flare or Flash (2942)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the customer smelled smoke when turning on the machine.No injury reported.A field engineer was dispatched to the site and found a transformer within the generator cabinet appeared melted.The generator cabinet and cables to the table were replaced.Once this was completed the system was working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFIRM PRONE BIOPSY SYSTEM, 3D
Common Device Name
BREAST BIOPSY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
MDR Report Key11803017
Report Number1220984-2021-00007
Device Sequence Number3659546
Product Code IZH
UDI-Device Identifier15420045506510
UDI-Public15420045506510
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K153486
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPBX-SYS-AFFIRM-3D
Device Catalogue NumberPBX-SYS-AFFIRM-3D
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2021
Initial Report FDA Received Date05/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-