| Model Number |
ZCB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Weight & ethnicity: information unknown/not provided.Date of event: exact date unknown/not provided.Best estimate date is between 1/14/2021-3/11/2021.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an intraocular lens (iol) was explanted from the patient's operative eye due to the patient experiencing decreased and blurry vision.Visual acuity pre-op: 20/30 -2 uncorrected.Visual acuity post-op: 20/25 uncorrected.There was no vitrectomy, incision enlargement, or sutures required.Another johnson & johnson lens (same model and higher diopter) was implanted as a replacement.The patient is very happy with the outcome.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9: device available for evaluation? yes.Returned to manufacturer on: jun 9, 2021.Section h3: device evaluated by manufacturer¿ yes.Device evaluation: the complaint lens was received in a specimen cup.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptics and that the lens was received cut in half, which is consistent with a lens that was handled during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.The search revealed that no similar complaint for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|