| Model Number |
3910-400-055 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material in the sterile packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: packaging compromised.Probable root cause: process uncontrolled environmental conditions.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material in the sterile packaging.
|
| |
|
Search Alerts/Recalls
|