• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. VISUALASE; LASER INSTRUMENT, SURGICAL, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. VISUALASE; LASER INSTRUMENT, SURGICAL, POWERED Back to Search Results
Model Number 9735560
Medical Device Problem Codes Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a thermal therapy system being used in a soft tissue ablation (neuro).It was reported that there were two issues that were discovered with the kit.The first issue was that the protective tube had a wavy form and markings in the bends.The second issue was the that the health care professional discovered a mark at 60 mm from the diffusion tip.The reported issue was found during the verification process.Troubleshooting was performed and the catheter was inspected and no damage was found.Light was applied to the fiber and there was no apparent diffusion at the 60 mm mark.As the marking was unknown it was decided to get a new kit to be on the safe side.This one also had this marking and had a different lot number so it was decided to proceed with the second fiber.There was less than an hour delay in the case.No impact on patient outcome.Additional information was received stating that the reported cause of the event is unknown.They received confirmation from senior laser ablation surgical consultant that the marking at 60mm was part of production.Specifically the mark is second fiber protection layer that ends.Inspection didn't show any damage to the catheter.The second one from a different lot will not be returned as there's nothing wrong with the fiber.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISUALASE
Common Device Name
LASER INSTRUMENT, SURGICAL, POWERED
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key11806035
Report Number1045254-2021-00236
Device Sequence Number17058633
Product Code GEX
UDI-Device Identifier00763000218188
UDI-Public00763000218188
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K053087
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/05/2023
Device Model Number9735560
Device Catalogue Number9735560
Device Lot Number0221456724
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2021
Initial Report FDA Received Date05/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age41 YR
Patient Weight90
-
-