Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultracor breast biopsy marker was returned for evaluation.The marker appeared to be have blood residue on the tip of the marker.The marker system was returned in a pre-deployed configuration as the pusher was pulled back and the trigger was in the locked firing position.It was noted there was a slight bend to the cannula notch.The wireform was not returned for evaluation.No functional testing performed due to the condition of the device.Therefore, the investigation is confirmed for bent needle as the needle found to be bent.However, the investigation is inconclusive for separation failure and difficult to remove as the wireform was not returned along with the device and the used condition of the device on patient can not be replicated.A definitive root cause for the alleged separation failure, difficult to remove and bent needle could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date: 02/2023).
|