Model Number 04.511.204.01
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that, the titanium matrix screw self tapping was removed from the patient and the self retaining would not engage in the screw.Procedure was completed successfully with fifteen (15) minutes surgical delay and no patient consequences.Concomitant devices reported: unknown matrixorthognathic plate (part# unknown, lot# unknown, quantity 1), unknown screwdriver (part# unknown, lot# unknown, quantity 1).This report is for one (1) 1.85mm ti matrix screw self-tapping/4mm.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A manufacturing record evaluation could not be performed as the lot number could not be determined from the available information.Also could not find any markings on the part.Visual inspection: the 1.85mm ti matrix screw self-tapping/4mm was returned and received at us customer quality (cq).Upon visual inspection, it was observed that the thread of the self-retaining screw is found stripped near proximal side of screw.The compliant condition of screw not able to self-retained may be confirmed due to threads of the screw being stripped.No other issues were identified with the returned device.Functional test: the functional test cannot be performed on the self-retaining screw as only the screw was returned and could not replicate the use of device through laboratory analysis.Dimensional inspection: the dimensional inspection could not be performed due to post manufacturing damage.Document/specification review: due to lot code being unknown, the date of manufacture cannot be confirmed and only the current revision of drawings were reviewed.Investigation conclusion the complaint condition was confirmed for the 1.85mm ti matrix screw self-tapping/4mm.During the investigation, no product design issues or discrepancies were observed.No manufacturing issues were noted during investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.85MM TI MATRIX SCREW SELF-TAPPING/4MM
Common Device Name
PLATE, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11806724
Report Number2939274-2021-02305
Device Sequence Number14327584
Product Code JEY
UDI-Device Identifier10887587023410
UDI-Public(01)10887587023410
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K083388
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.511.204.01
Device Catalogue Number04.511.204.01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2021
Initial Date Received by Manufacturer 04/15/2021
Supplement Date Received by Manufacturer06/11/2021
Initial Report FDA Received Date05/11/2021
Supplement Report FDA Received Date06/11/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MATRIXMDF SCRDRVR BLD HEX CPL/S-RETN/76; UNK - PLATES: MATRIXORTHOGNATHIC
-
-