|
SMITH & NEPHEW, INC. 12.5MM ENTRY REAMER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
| Model Number |
71631116 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during service, a 12.5mm entry reamer was found to be chipped and dulled.No case involved; therefore, there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|