MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ACCU-PASS STR SHUTTLE 45 DEG RGHT CRVE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ACCU-PASS STR SHUTTLE 45 DEG RGHT CRVE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 7210424
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during an acl procedure, at the moment of passing the suture through the accu-pass through the labrum, after the anchor was placed, the device broke in the part where the curve begins to the right.Part was successfully removed from the patient.Backup device was available to complete the procedure.No significant delay and no patient complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was a relationship found between the subject device and the reported incident.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A review of the customer provided image reveals a monofilament is in the device and the distal tip of the needle has become separated from the shaft.The complaint was confirmed.Factors, which could have contributed to the complaint event, include an application of unintended inappropriate or excessive force to the device.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-PASS STR SHUTTLE 45 DEG RGHT CRVE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key11808974
Report Number1219602-2021-01093
Device Sequence Number3909917
Product Code LXH
UDI-Device Identifier03596010534453
UDI-Public03596010534453
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/17/2023
Device Model Number7210424
Device Catalogue Number7210424
Device Lot Number2057177
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/22/2021
Supplement Date Received by Manufacturer06/15/2021
Initial Report FDA Received Date05/11/2021
Supplement Report FDA Received Date06/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-