Model Number 530.792
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.As of this date, the device has not been returned for evaluation.Therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Event or Problem Description
It was reported that the hudson drilling attachment device did not keep the attachment spinning once there was any resistance applied.The reporter stated that the device had never been used in a surgical setting with a live patient.The event was not related to surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device had a broken driveshaft and a damaged component.It was further determined that the device failed pretest for check for untrue running.The assignable root cause was determined to be due to component failure from normal wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON DRILLNG ATTACH BATT POWER LINE
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key11810200
Report Number8030965-2021-03787
Device Sequence Number14394568
Product Code HWE
UDI-Device Identifier10886982240279
UDI-Public10886982240279
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number530.792
Device Catalogue Number530.792
Device Lot Number8804821
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2021
Supplement Date Received by Manufacturer05/14/2021
Initial Report FDA Received Date05/11/2021
Supplement Report FDA Received Date05/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-