| Model Number |
KD-VC611Q-07301S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device was returned to olympus¿s local distributor, but not returned to omsc.Therefore the exact cause of the reported event could not be conclusively determined at this time.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
|
| |
|
Event or Problem Description
|
|
Olympus medical systems corp.(omsc) was informed by the health professional that during an endoscopic sphincterotomy (est) using the subject device, the cutting wire was broken.There were no missing parts.The intended procedure was completed with another device.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This is a supplemental report to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The cutting wire was broken.Investigation was carried out to confirm the broken portion.The coated portion of the cutting wire was torn, and the broken portion was scorched and melted.The outer diameter of the cutting wire was measured.The result indicated no abnormalities.The length of the coated portion of the cutting wire, and the cutting wire itself presented no abnormalities.There were no missing parts in the subject device.Other abnormalities that could lead to the breakage of the cutting wire were not confirmed.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, omsc presumes that the event occurred due to the following occurrence mechanism.1.The cutting wire at a torn area of the coated portion came into contact with the distal end of the endoscope while the forceps elevator was raised.2.The output was activated in state of ¿1¿ description, and the cutting wire became hot instantly.This caused the cutting wire to break.It has been confirmed that the tear of the coated portion of the cutting wire could duplicate by the following mechanism.1.The forceps elevator of the endoscope was raised.2.When the cutting wire deflects, the coated portion of the cutting wire and the metal part of the distal end of the endoscope come into contact.(see fig.2) 3.In state of description ¿2¿, the cutting wire was moved back and forth.This caused the coated portion of the cutting wire to tear.(see fig.3) the slider was pushed more than needed.This caused the cutting wire to deflect.The above device handling has warned in the instruction manual.
|
| |
|
Search Alerts/Recalls
|