• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, UNIV ENDS,; SYSTEM/DEVICE, PHARMACY COMPOUNDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, UNIV ENDS,; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number H93821
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter phone no.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that a fluid transfer tube set was leaking with unspecified medicine.The leak was coming from the white area of the set that fits into the pump.A new set was replaced and the issue resolved.This issue was identified during priming.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
H4: device manufactured between july 06, 2020 to july 08, 2020.H10: the device was received for evaluation.Unaided eye visual inspection was done which observed a hole at a torn section of the silicone tubing of the tube set.A functional testing was performed which revealed a leak at the silicone tubing area where the hole was located.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUBE SET, STD VOL, UNIV ENDS,
Common Device Name
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11811896
Report Number1416980-2021-02837
Device Sequence Number3249168
Product Code NEP
UDI-Device Identifier00085412476520
UDI-Public(01)00085412476520
Combination Product (Y/N)Y
Initial Reporter CountryCO
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2023
Device Catalogue NumberH93821
Device Lot Number60245800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2021
Initial Date Received by Manufacturer 04/16/2021
Supplement Date Received by Manufacturer06/04/2021
Initial Report FDA Received Date05/12/2021
Supplement Report FDA Received Date06/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-