MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE TUBING SET; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE TUBING SET; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10031
Medical Device Problem Code Inability to Irrigate (1337)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a tka surgery the irrigation didn¿t work during milling.The irrigation tubing was able to prime outside of the pump during milling, but once it was placed the tubing back within the pump there was no irrigation.The procedure was completed, without delay, using a bulb irrigation.It is unknown if this was a tubing or a console issue.No patient injuries and no other complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the ri tubing set (us), rob10031, lot 50891743 used for treatment, was not returned for evaluation; thus, a visual and functional evaluation could not be performed, and a relationship between the reported event and the device could not be confirmed.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.A capa, hhe/pra, field action review was not completed.The product was not returned and no evidence was made available to link the complaint to a capa, hhe/pra, field action.Although the reported problem was not confirmed, a factor that may have contributed to the reported symptom may be associated with a mechanical component failure.Refer to real intelligence cori for knee arthroplasty user manual to the preparing the irrigation tubing section that provides guidelines for setting up the irrigation tube.This failure is an identified failure mode within the risk assessment.Per complaint details from the united states it was reported that during a cori assisted tka surgery, the irrigation didn't work during milling.The irrigation tubing was able to prime outside of the pump during milling, but once it was placed back in the pump there was no irrigation.The procedure was completed, without delay, using a bulb irrigation.No patient injuries and no other complications were reported.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE TUBING SET
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11813031
Report Number3010266064-2021-00360
Device Sequence Number3766581
Product Code OLO
UDI-Device Identifier00885556757468
UDI-Public00885556757468
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberROB10031
Device Catalogue NumberROB10031
Device Lot Number50891743
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2021
Supplement Date Received by Manufacturer12/20/2021
Initial Report FDA Received Date05/12/2021
Supplement Report FDA Received Date12/20/2021
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-