| Catalog Number |
53035KA2 |
| Medical Device Problem Codes |
Material Rupture (1546); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a peripheral vascular procedure, the catheter tip detached outside the patient.The physician had successfully acquired retrograde arterial access, and the catheter was introduced over a stiff guidewire, the lesion was crossed, and contrast was injected to collect selective angiograms.The catheter was removed from the patient and placed back on the sterile field.At a later point within the procedure, the clinician decided to use the 5f ka2 catheter again.When an attempt was made to reload the catheter over the back end of the guidewire, the tip detached and was dangling from the catheter shaft.A new catheter was opened and used to complete the procedure.No patient injury to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned for evaluation.The complaint is confirmed.The root cause is attributed to the manufacturing process.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|
Search Alerts/Recalls
|