• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number BI70002000
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No analysis was completed at this time.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that mobile viewing station (mvs) wont power on.The site reported that while they were storing the imaging system, the system would not charge up or turn on.There was no patient present.
 
Additional Manufacturer Narrative
Additional information: b5 updated with additional information received.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received.The system was plugged in while it was in storage.It was found the incoming 230v fuses had blown.Once the fuses were replaced, and the lower circuit breaker turned on, the system powered up fine.
 
Additional Manufacturer Narrative
Information references the main component of the system.Other relevant device(s) are: product id: bi71000522, serial/lot #: none (x2) ,a representative attended the site and inspected the mobile viewing station.The incoming fuses had blown.The fuses were replaced and the system was ready for use.Codes b01, c02 and d02 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
MDR Report Key11816837
Report Number3004785967-2021-00617
Device Sequence Number3223221
Product Code OWB
UDI-Device Identifier00763000134518
UDI-Public00763000134518
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K200074
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2021
Supplement Date Received by Manufacturer05/17/2021
05/17/2021
Initial Report FDA Received Date05/13/2021
Supplement Report FDA Received Date05/18/2021
05/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-