Model Number 71070410
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during delivery inspection, it was found that the tsf torque wrench 10mm was rusty.As this was found during inspection, there was no patient involvement.
 
Additional Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was not returned for evaluation.Per the photo attached the failure was confirmed.The device was found to be discolored, rendering the device inoperable.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TSF TORQUE WRENCH 10MM
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11816886
Report Number1020279-2021-04157
Device Sequence Number19620248
Product Code JWH
UDI-Device Identifier03596010462336
UDI-Public03596010462336
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K121443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71070410
Device Catalogue Number71070410
Device Lot Number19DK0065
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/22/2021
Supplement Date Received by Manufacturer06/11/2021
Initial Report FDA Received Date05/13/2021
Supplement Report FDA Received Date06/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-