SMITH & NEPHEW, INC. MI TRIAL FEM HEAD 36 +4; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Model Number |
71369722 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a thr, a mi trial fem head 36 +4 was observed to have one of the four locking tabs broken.Tabs are used to secure the trial head to the reunion of the stem or the trial neck.Instrument outside the patient.The procedure was able to be completed with the same device.Surgery was not delayed.Patient was not harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device, intended for use in treatment, was not returned for evaluation.Per the photo attached the failure was confirmed.The tab is broken, rendering the device inoperable.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|