• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® SILASTIC® FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® SILASTIC® FOLEY CATHETER Back to Search Results
Model Number 33616
Medical Device Problem Codes Component Misassembled (4004); Missing Information (4053)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that a piece of the foley catheter arrived incomplete because it lacked the identification color.
 
Event or Problem Description
It was reported that a piece of the foley catheter arrived incomplete because it lacked the identification color.
 
Additional Manufacturer Narrative
The reported event was confirmed manufacturing related.1 sample was confirmed to exhibit the reported failure.The device had not met specifications.The product was not used for patient treatment the product caused the reported failure.Visual evaluation of the returned sample noted one unopened (within original packaging), unused silastic foley.Visual inspection of the sample noted that the inflation cap was missing.This does not meet the specification "the cap must be present, the height and position must be correct." a potential root cause for this failure could be operator error.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labelling could not have prevented the reported failure.The actual/suspected device was evaluated.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing-related.One sample was confirmed to exhibit the reported failure.The device had not met specifications.The product was not used for patient treatment the product caused the reported failure.Visual evaluation of the returned sample noted one unopened (within original packaging), unused silastic foley.Visual inspection of the sample noted that the inflation cap was missing.This does not meet the specification as "the cap must be present, the height and position must be correct.".A potential root cause for this failure could be operator error.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.Labeling review was not performed because labelling could not have prevented the reported failure.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that a piece of the foley catheter arrived incomplete because it lacked the identification color.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® SILASTIC® FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11820057
Report Number1018233-2021-02763
Device Sequence Number14394452
Product Code EZL
UDI-Device Identifier00801741020063
UDI-Public(01)00801741020063
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K951103
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model Number33616
Device Catalogue Number33616
Device Lot NumberNGEW1836
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/20/2021
Supplement Date Received by Manufacturer12/08/2021
07/19/2022
Initial Report FDA Received Date05/13/2021
Supplement Report FDA Received Date12/13/2021
07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-