|
Model Number 0998-00-3023-53 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported issue.
The fse discovered that someone had switched off the charging unit.
A complete system checkout and reset the battery maintenance required message was performed.
The fse then performed all functional and safety checks to meet factory specifications.
Unit passed all functional and safety test per factory specifications.
The iabp was charging the batteries once the switch was turned back on, and was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during a routine check, the cs300 intra-aortic balloon pump (iabp) was not charging the batteries.
There was no patient involvement, and no adverse event reported.
|
|
Search Alerts/Recalls
|
|
|