Model Number 0998-00-3023-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported issue.The fse discovered that someone had switched off the charging unit.A complete system checkout and reset the battery maintenance required message was performed.The fse then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was charging the batteries once the switch was turned back on, and was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that during a routine check, the cs300 intra-aortic balloon pump (iabp) was not charging the batteries.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Analysis of production: (3331) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109) the review of the historical data indicates that no other similar complaint was reported for the same serial number and reported failure mode.Trend analysis: (4110) the overall 24 month product complaint trend data for the period (may-2019 through apr-2021) was reviewed.There were no triggers identified for the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
MDR Report Key11820750
MDR Text Key250846665
Report Number2249723-2021-01006
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/21/2021
Initial Date FDA Received05/13/2021
Supplement Dates Manufacturer Received07/29/2021
Supplement Dates FDA Received07/29/2021
Patient Sequence Number1
-
-