It was reported by the sales rep in (b)(6) that post-operatively to an arthroscopic shoulder stabilization procedure on (b)(6) 2021, it was observed that the suture knot on the 5.5mm healix advance¿ br anchor with dynacord¿ suture (blue, white/blue/green, white/black) untied that the patient needed a revision surgery.The status of the patient post-revision surgery was unknown.No additional information was provided.
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary the complaint device is not being returned, it was implanted in the patient, therefore unavailable for a physical evaluation.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.Given that lot number was not provided, the manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|