• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND AFFINITI 30; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND AFFINITI 30; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 989605457591
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The philips field service engineer performed a system inspection and determined that the device does not lock when rotated.The service engineer replaced the control panel articulating arm to correct the issue.An investigation is underway to determine the root cause of the issue.Results of the investigation will be included in a follow up report upon its completion.
 
Event or Problem Description
The customer reported an incident where the swivel mechanism of their affiniti ultrasound system¿s control panel would not lock into position while transporting the unit within the user facility.No patient or user was harmed as a result of the issue.The ultrasound system has been returned to service with no additional issues reported.
 
Additional Manufacturer Narrative
A thorough investigation was performed to identify the root cause of the reported issue.However, due to insufficient information and part returns, the engineering team was unable to determine the root cause of this failure.The control panel articulation arm was replaced to resolve the issue.Actions are underway to ensure the necessary information and parts are received to further investigate if this issue does recur.
 
Event or Problem Description
The customer reported an incident where the swivel mechanism of their affiniti ultrasound system¿s control panel would not lock into position while transporting the unit within the user facility.No patient or user was harmed as a result of the issue.The ultrasound system has been returned to service with no additional issues reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITI 30
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
eric nail
22100 bothell everett highway
bothell, WA 98021
MDR Report Key11821868
Report Number3019216-2021-10059
Device Sequence Number19791284
Product Code IYN
UDI-Device Identifier00884838074187
UDI-Public00884838074187
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K160807
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number989605457591
Device Catalogue Number795218
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date04/19/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/13/2021
Supplement Report FDA Received Date03/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/08/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-