Model Number 0998-UC-3013-53 |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate.Upon switching the unit on, the unit showed "low battery message." further testing revealed that the on/off switch on the power supply was in off position.After switching the power supply to on, the battery began to charge and the low battery message disappeared.The fse observed the unit with a trainer and balloon connected and the unit worked fine.The fse recommended that the customer should keep the unit charging.The fse then performed a full pm with calibration, functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.(b)(6).
|
|
Event Description
|
It was reported that during use on a patient, the cs100 intra-aortic balloon pump (iabp) was not working.No patient harm, serious injury or adverse event was reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Analysis of production: (3331) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported serial number.Based on the dhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109) the review of the historical data indicates that no other similar complaint was reported for the same serial number and reported failure mode.Trend analysis: (4110) the overall 24 month product complaint trend data for the period (may 2019 through apr 2021) was reviewed.There were no triggers identified for the review period.
|
|
Search Alerts/Recalls
|
|
|