Model Number 0998-UC-3013-53
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate.Upon switching the unit on, the unit showed "low battery message." further testing revealed that the on/off switch on the power supply was in off position.After switching the power supply to on, the battery began to charge and the low battery message disappeared.The fse observed the unit with a trainer and balloon connected and the unit worked fine.The fse recommended that the customer should keep the unit charging.The fse then performed a full pm with calibration, functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.(b)(6).
 
Event Description
It was reported that during use on a patient, the cs100 intra-aortic balloon pump (iabp) was not working.No patient harm, serious injury or adverse event was reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Analysis of production: (3331) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported serial number.Based on the dhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109) the review of the historical data indicates that no other similar complaint was reported for the same serial number and reported failure mode.Trend analysis: (4110) the overall 24 month product complaint trend data for the period (may 2019 through apr 2021) was reviewed.There were no triggers identified for the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
MDR Report Key11822043
MDR Text Key251198558
Report Number2249723-2021-01009
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107745
UDI-Public10607567107745
Combination Product (y/n)N
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-UC-3013-53
Device Catalogue Number0998-UC-3013-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/29/2021
Patient Sequence Number1
Treatment
LINEAR.
-
-