| Model Number |
16021 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
04/13/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history records found no manufacturing, processing or design related irregularities.The instrument was found to be properly manufactured and released in accordance with the device master record.Instrument (16021) was used to supplement a final driver (16022).The 16021 universal cross connector inserter is not designed for the input loads of the final driver sequence.
|
| |
|
Event or Problem Description
|
|
The torque handle did not unlock the cross connector so the inserter was used but it broke inside the cross connector.No injury to the patient but a piece of the inserter driver was retained in the patient.
|
| |
|