| Model Number |
CRPLUS |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to physio-control for evaluation.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device displayed all three icons (charge-pak, attention, and service wrench).The illumination of all three icons indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use reported for this event.
|
| |
|