• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number 10371
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 04/20/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that thrombosis occurred.A left atrial appendage (laa) closure procedure was performed and a 27mm watchman laa closure device was implanted.The 45-day transesophageal echo (tee) was unremarkable.The patient had some neurological changes which prompted a tee in may 2019 which showed a device related thrombus.The patient was put back on oral anticoagulation (eliquis) and the clot resolved.The patient came in last week with new neurological changes.A tee was performed and a new clot had formed on the face of the device measuring 16x5mm.The patient was put back on eliquis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM
Common Device Name
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11822536
Report Number2134265-2021-06320
Device Sequence Number1061029
Product Code NGV
UDI-Device Identifier08714729838241
UDI-Public08714729838241
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/20/2020
Device Model Number10371
Device Catalogue Number10371
Device Lot Number0021419170
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/26/2021
Initial Report FDA Received Date05/13/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-