• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941838350
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Codes Pain (1994); Thrombosis/Thrombus (4440)
Date of Event 04/21/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that stent thrombosis occurred.A 3.50 x 38mm synergy xd drug-eluting stent was implanted after non-boston scientific(bsc) intravascular lithotripsy (ivl) as well as other non-bsc revascularization techniques were performed.During the procedure, the activated clotting time(act) was therapeutic, however; the subject came back to the lab with pain a few hours later.Intravascular ultrasound(ivus) verified that the stent had thrombus and it was also undersized.The stent was then post dilated to properly size it within the lesion.There were no further patient complications reported and the patient was doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11822649
Report Number2134265-2021-05820
Device Sequence Number1061031
Product Code NIQ
UDI-Device Identifier08714729981091
UDI-Public08714729981091
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/05/2023
Device Model NumberH7493941838350
Device Lot Number0026648278
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/21/2021
Initial Report FDA Received Date05/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-