| Model Number |
1188410105 |
| Medical Device Problem Code |
Fogging (1253)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a foggy image.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: foggy.Probable root cause: cables; connectors; digital board; power supply; filter/fuse; ac inlet board; main board; transition board; auxiliary board; intermittence between transition board and ch connector; fan/insufficient cooling; software; camera head (sensor, sensor cable); coupler (dust on optics, scratched optics, broken/shifted lenses, focusing mechanism, endoclamp); extension cable; intermittence while using rf devices (ex: valleylab); problem toggling light source from camera head; connected monitors; leaking; manufacturing/service nonconformity; use error.The reported failure mode will be monitored for future reoccurrence.H3 other text: 81.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a foggy image.
|
| |
|
Search Alerts/Recalls
|