• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTSINE TECHNOLOGIES LTD PACKAGE,350P,PP01,EN,350-STR-US-10; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTSINE TECHNOLOGIES LTD PACKAGE,350P,PP01,EN,350-STR-US-10; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number SAM 350P
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2021
Type of Reportable Event Malfunction
Event or Problem Description
Low battery warning and beep with in date pad-pak.No patient involvement.
 
Additional Manufacturer Narrative
Heartsine's investigation of the device confirmed the reported fault, as the returned pad-pak was found to be severely depleted.Upon receipt, the device could not be powered on.These faults were attributed to moisture and corrosion within the device on multiple components and circuits.This had led to a constantly illuminated green status indicator alongside a flashing red status indicator on receipt at heartsine, which resulted in a high current drain and the subsequent depletion of the returned pad-pak.Inability to power the device on could result in adverse consequences, should it be used in a patient event.
 
Event or Problem Description
Low battery warning and beep with in date pad-pak.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGE,350P,PP01,EN,350-STR-US-10
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI  BT3 9ED
MDR Report Key11824258
Report Number3004123209-2021-00156
Device Sequence Number11863085
Product Code MKJ
UDI-Device IdentifierM727SAM350P
UDI-PublicM727SAM350P
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
P160008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberSAM 350P
Device Catalogue Number350-STR-US-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2021
Initial Date Received by Manufacturer 05/04/2021
Supplement Date Received by Manufacturer08/17/2021
Initial Report FDA Received Date05/14/2021
Supplement Report FDA Received Date08/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-