According to the reporter, the catheter was placed in patient and flushed without issues.The patient went to dialysis for (hemodialysis) hd treatment and the catheter began to leak from the venous extension tubing underneath the clamp.A new catheter of the same type was placed.The new catheter was used successfully.The procedure was completed.Nothing unusual observed prior to use.The catheter was not repaired, and there was no luer adapter issue.No other intervention/treatment required.There was no patient injury.
|
|
Additional information: d9, g1(manufacturer name, mfr contact first name, last name, street1, mfr city, region, postal code, email, phone number), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that there was a cut/hole/disconnection in the extension tube which led to a leak.It was reported that there was a leak or hole on the extension tube.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when contact is made with a sharp surgical instrument during the clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: exercise caution when using sharp instruments near the catheter.Catheter tubing can tear when subjected to nicks, excessive force, or rough edges.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|