| Model Number |
2054 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been returned to the local facility but has not yet been received at the main office for evaluation.Once returned and investigated, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported the sensor stops working suddenly.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: other, other text: the returned device was evaluated.External visual inspection found no damage.The sensor failed continuity testing due to an open in the cable on the detector circuit along the length of the cable.When connected to a monitor a "sensor off patient" error message was provided.
|
| |
|
Event or Problem Description
|
|
The customer reported the sensor stops working suddenly.No patient impact or consequences were reported.
|
| |
|