• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 32MM X 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 32MM X 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71339550
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Muscular Rigidity (1968); Loss of Range of Motion (2032); Arthralgia (2355)
Date of Event 04/29/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after daa tha surgery, the patient was in pain from the second week and onwards, felt tight across the front hip, did not had a good range of movement and the leg on the operative side was shorter.A revision surgery was performed on(b)(6) 2021 and all the components were explanted.The patient outcome is unknown.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, the provided x-rays were reviewed; however, the events related to the complaint are unclear.On the provided images only the right side appears to have an implant.The left side appears to be planning pre-op for a left sided procedure; therefore we can¿t make any conclusions from the left sided pre-op planning.Smith and nephew has not received the explanted devices and it is noted no additional information is available.If additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes could be alignment, fit/size of device used or wear.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 0 DEG XLPE ACET LNR 32MM X 50MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11826657
Report Number1020279-2021-04333
Device Sequence Number3535259
Product Code JDI
UDI-Device Identifier03596010598806
UDI-Public03596010598806
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71339550
Device Catalogue Number71339550
Device Lot Number20GM02592
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/29/2021
Supplement Date Received by Manufacturer08/26/2021
Initial Report FDA Received Date05/14/2021
Supplement Report FDA Received Date08/30/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-